번역 버튼을 눌러보세요!
Reticulocyte Counting of the Rat Blood Using Flowcytometry in Comparison with Manual Methods Flowcytometry를 이용한 랫드 망상적혈구의 측정
논문 기본 정보
- 저자정보
- 저널정보
- 한국실험동물학회 Laboratory Animal Research THE KOREAN JOURNAL OF LABORATORY ANIMAL SCIENCE Vol.15 No.3
- 발행연도
- 1999.12
- 수록면
- 257 - 260 (4page)
Reticulocyte Counting of the Rat Blood Using Flowcytometry in Comparison with Manual Methods
초록·키워드
The reticulocyte counting is an important parameter in assessing of subject with anemia in laboratory animals and humans. The manual counting method is thought to have some limitation with low precision and heavy work load. Recently fluorescent activated cytometric (FACS) analysis of reticulocytes by thiazol orange (TO) has been introduced in human hematology to improve the precision and the speed of measurement. The reticulocyte of rat blood was analyzed by manual method and FACS, and compared. A good correlation (r = 0.897, slope = 10.03, intercept = -3.50) between manual count and FACS determination was observed. The methodology of FACS was accurate and cost-effective, and showed a high correlation with manual techniques.
본문·목차
인공지능 문자 인식 모델을 통해 추출된 텍스트로, 일부 오타나 오류가 포함될 수 있으나 지속적으로 개선 중입니다.
오류를 발견하셨다면 해당 부분을 드래그한 후 ' 를 통해 신고해주세요.
오류를 발견하셨다면 해당 부분을 드래그한 후 ' 를 통해 신고해주세요.
최근 본 자료 전체보기
UCI(KEPA) : I410-ECN-0101-2009-510-016326096