High speed and high resolution UHPLC with 140 MPa maximum operating pressure by using a sub-2μm packed 250mm-length column
High speed and high resolution UHPLC with 140 MPa maximum operating pressure by using a sub-2μm packed 250mm-length column
초록·키워드
Current commercially available ultra high-performance liquid chromatography (UHPLC) works within aultra high operating pressure (max. 100 MPa or more) and it allows utilizing sub-2 μm particle columns. UHPLC analysis offers a significant reduction in analysis time without degradation of peak resolution. It also allows an increase in sensitivity and high resolutionanalysis. Pressure dropΔP (MPa) is proportional to flow rate u (mm/s), column lengthL (mm), and mobile phase viscosityη (Pa·s) and inversely proportional to column permeability Kᵥ (m²) (see eq.1); On this equation, you can find the extension of maximum working pressure provides to enhance those three proportional parameters. Expansion of flow rate leads to analytical time shorter, which leads to high-speed analysis. Expansion of column length means improvement of the theoretical plate, which means high-resolution analysis. And extension of the mobile ... 전체 초록 보기