    Hemodynamic performance of a polyurethane heart valve prosthesis was evaluated in comparison with that of Björk-Shiley Monostrut mechanical valve in steady flow representing the systolic peak flow phase. Pressure losses through the valves were obtained from the streamwise pressure distributions downstream of the valves. Unsteady and turbulent flow field distal to the heart valve prostheses were investigated using PIV (Particle Image Velocimetry) which can measure the full-field velocity instantaneously and noninvasively. By examining the velocity and Reynolds shear stress fields downstream of the polyurethane heart valve, it is known that there is a large recirculation region near the valve and high shear stress regions exist at the interface between strong axial jet flows along the wall and vortical flows in the central area. The possibilities of vascular complications, such as the thrombus formation and red blood cell damage, could be predicted from the overall view of the velocity and stress fields.