저자정보
최병욱 (연세대학교 의과대학 진단방사선과교실, 방사선의과학연구소)
저널정보
대한영상의학회 대한방사선의학회지 대한방사선의학회지 제40권 제2호
발행연도
1999.1
수록면
317 - 324 (8page)

이용수

표지
📌
연구주제
📖
연구배경
🔬
연구방법
🏆
연구결과
AI에게 요청하기
추천
검색

초록· 키워드

오류제보하기
Purpose : To determine the effect of anglulation between aorta the and renal artery on signal loss in theproximal renal artery, as seen on magnetic resonance angiography by phantom study using a pulsatile flow model.Materials and Methods : Three phantoms of aorta and renal artery with angulation of 90。, 60。, and 30。 wereobtained. Pulsatile recirculating flow (44%W/W glycerin, 60bpm) was used for MR angiography. First, axial 3D-TOFimages were obtained and reconstructed. MIP images were analyzed for the presence, area, and location of signalloss. 2D-PC images were obtained perpendicularly to the renal artery at a distance of 0, 4, 8 and 12mm from theostium. To calculate mean signal intensity of the renal artery, a ROI was drawn on 2D-PC images. To correlatesignal loss in 3D-TOF images with signal decrease in 2D-PC, we analyzed changes in signal intensity during onepulse cycle according to change of angulation and distance from the ostium of the renal artery by the calculatedvalues of relative signal decrease and ratio of signal decrease. Results : A signal loss was observed up to 4mmfrom the ostium of the renal artery only in the case of the 90。 phantom. Because the signal intensity measured inthe 2D-PC image of the 90。 phantom was higher than that of the 60。 phantom the signal loss observed in the3D-TOF images of the 90。 phantom could not be explained by the magnitude of measured signal intensity alone.Relative signal decrease only at a distance of 0 and 4mm in the 90。 phantom was evenly increased through a pulsecycle and the ratio of signal decrease at the same location was more than 50%. In contrast to the results of the90。 phantom, those of 60。 and 30。 showed decreased of signal intensity mainly during the diastolic phase.Conclusion : Signal loss should become apparent at a certain angle between 60。 and 90。. Decreased signalintensity causing signal loss in 3D-TOF was maintained throughout the systolic and diastolic phase of a pulsatilecycle and correlated with the ratio of signal decrease.

목차

등록된 정보가 없습니다.

참고문헌 (0)

참고문헌 신청

함께 읽어보면 좋을 논문

논문 유사도에 따라 DBpia 가 추천하는 논문입니다. 함께 보면 좋을 연관 논문을 확인해보세요!

이 논문의 저자 정보

최근 본 자료

전체보기

댓글(0)

0