메뉴 건너뛰기

추천
검색

논문 기본 정보

자료유형
학술저널
저자정보
(서울의대 방사선과학교실)
저널정보
대한영상의학회 대한영상의학회지 대한방사선의학회지 제23권 제1호
발행연도
수록면
113 - 116 (4page)

이용수

표지
📌
연구주제
📖
연구배경
🔬
연구방법
이 논문의 연구방법이 궁금하신가요?
🏆
연구결과
이 논문의 연구결과가 궁금하신가요?
AI에게 요청하기
추천
검색

초록· 키워드

CT scanogram is a new technique for measuring leg length discrepancy and has many advantages as compared with conventional spot scanogram. 30 dried bones wetre measured by direct measurement, spot scanogram and CT scanogram. Accuracy of CT scanogram was compared with spot scanogram. For evaluation of systemic error of CT, steel rulers and dried bones were measured at two different window settings. Decentering error in spot scanogram was measured by using dried bones and 180 spots of 30 clinical cases of Bell-Thompson studies were reviewed for evaluation of incidence and degree of decentering. The results of measurements of CT and spot scanogram in ideal position and technique were equivalent to those of direct measurements. But in spot scanogram, significant errors were neoted in flexion status and decentered film. 54 spots of 30 clinical cases were decentered more than 1cm. In CT scanogram, decentering error was completely excluded and shortening due to flexion deformity was vercomed by lateral scanogram. CT scanogram is a simple and accurate method for leg length measurement. Thus we recommend CT scanogram as a routine use for evaluation of leg length discrepancy.
상세정보 수정요청해당 페이지 내 제목·저자·목차·페이지
정보가 잘못된 경우 알려주세요!

목차

등록된 정보가 없습니다.

참고문헌

참고문헌 신청

최근 본 자료

전체보기