Background: The main goal of experiments is to compare various operational and technical characteristics of D-Shuttle semiconductor personal dosimeters of the Japanese company "Chiyoda Technol Corporation" and Harshaw thermoluminescent dosimeters (TLD) manufactured by "Thermo Fisher Scientific" and DTL-02 of the Russian Research and Production Enterprise (RPE) "Doza" by their occupational and calibration exposure at various dose equivalents from 0.5 to 20 mSv of gamma-radiation. Materials and Methods: Besides dosimeters DTL-02, D-Shuttle and Harshaw TLD, there were also used: (1) the primary reference radionuclide source Hopewell Designs IAEA: G10-1-12 with $^{137}Cs$ isotope (an error is not more than 6% and activity is 20 Ci), and (2) the verification device UPGD-2M of RPE "Doza" and installed in the National Center for Expertise and Certification of the Republic of Kazakhstan (Kapchagai, the National Center for Expertise and Certification). Results and Discussion: The main results of researches are the following: (1) TLDs for Harshaw 6600 and DVG-02TM have an approximately equal measurement accuracy of the individual dose equivalents in the range from 0.5 to 20 mSv of gamma-radiation. (2) Advantages of dosimeters for Harshaw 6600 are due to the high measurement productivity and opportunity to indicate the dose on the skin $H_p$(0.07). Advantages of DVG-02TM consist of operation simplicity and lower cost than of Harshaw 6600. (3) D-Shuttles are convenient for use in the current and the operational monitoring of ionizing radiation. Measurement accuracy and 10% linearity of measurements are ensured when D-Shuttle is irradiated with dose equivalents below 1 mSv at the equivalent dose rate not higher than $3mSv{\cdot}hr^{-1}$. This allows using D-Shuttle at a routine technological activity. Conclusion: The obtained results of experiments demonstrate advantages and disadvantages of D-Shuttle semiconductor dosimeters in comparison with two TLD systems of DVG-02TM and Harshaw 6600.

목차

등록된 정보가 없습니다.

참고문헌 (7)

참고문헌 신청

함께 읽어보면 좋을 논문

논문 유사도에 따라 DBpia 가 추천하는 논문입니다. 함께 보면 좋을 연관 논문을 확인해보세요!

이 논문의 저자 정보

최근 본 자료

전체보기

댓글(0)

0