(National Research Laboratory for Bioavailability Control, College of Pharmacy, Kangwon National University) (National Research Laboratory for Bioavailability Control, College of Pharmacy, Kangwon National University) (National Research Laboratory for Bioavailability Control, College of Pharmacy, Kangwon National University) (College of Agricultural Life Science, Seoul National University) (National Research Laboratory for Bioavailability Control, College of Pharmacy, Kangwon National University)
저널정보
대한약학회 Archives of pharmacal research : a publication of the Pharmaceutical Society of Korea Archives of pharmacal research : a publication of the Pharmaceutical Society of Korea 제30권 제8호
오류 신고하기

피인용 0

검색

    초록·키워드

    This study was performed in order to validate an effective high performance liquid chromatograpy (HPLC) method to determine L-carnitine in biological samples such as plasma, milk and muscle in cows. An L-carnitine derivative for fluorescence absorption was synthesized with 1-aminoanthracene (16 mg/mL in acetone) and 1-ethyl-3-(3-dimethylaminopropyl) carbodiimide hydrochloride (EDAC: 160 mg/mL in 0.01 M $NaH_2PO_4$ buffer) as a precolumn fluorescent derivative reagent. ${\gamma}-Butyrobetaine$ HCI was used as an internal standard. A reversed-phase column with fluorescence detection at the excitation and emission wavelengths of 248 and 418 nm were used. The mobile phase consisted of 30% acetonitrile with 0.1 M ammonium acetate in water (pH 3.5) adjusted with acetic acid and delivered at a flow rate of 1.5 mL/min. The L-carnitine concentration in plasma, milk and muscle samples of cows after oral feeding with 24 g L-carnitine/day for 2 months was then determined. All biological samples were deproteinated by barium hydroxide and zinc sulfate heptahydrate before the derivative reaction. Blank cow plasma was dialyzed using cellulose membrane for standard calibration. The calibration curve showed good linearity ($r^2$ >0.999) over the concentration range of 50 to 5000 ng/mL. The precision and accuracy were also satisfactory with less than 15% intra- and inter- day coefficiency of variations. The peaks of L-carnitine and internal standard in HPLC chromatography were successfully separated in plasma, milk and muscle samples of cows. The current derivatization method of L-carnitine for fluorescence detection was simple and adequately sensitive and could be applied to determine L-carnitine in biological samples.

    최근 본 자료 전체보기