논문 기본 정보

자료유형
학술저널
저자정보
Vasilios Alevizakos (Department of CAD/CAM and Digital Technologies in Dentistry Danube Private University Krems Austr) Richard Mosch (Department of CAD/CAM and Digital Technologies in Dentistry Danube Private University Krems Austr) Gergo Mitov (Department of Prosthodontics and Biomaterials Danube Private University Krems Austria) Ahmed Othman (Department of CAD/CAM and Digital Technologies in Dentistry Danube Private University Krems Austr) Constantin von See (Department of CAD/CAM and Digital Technologies in Dentistry Danube Private University Krems Austr)
저널정보
대한치과보철학회 The Journal of Advanced of Prosthodontics The Journal of Advanced of Prosthodontics 제13권 제3호
발행연도
2021.1
수록면
152 - 159 (8page)

이용수

표지
📌
연구주제
📖
연구배경
🔬
연구방법
🏆
연구결과
AI에게 요청하기
추천
검색

초록· 키워드

오류제보하기
Purpose. The aim of this study was to investigate to what extent cyclic load affects the screwless implant-abutment connection for Morse taper dental implants. Materials and Methods. 16 implants (SICvantage max) and 16 abutments (Swiss Cross) were used. The screwless implant-abutment connection was subjected to 10,000 cycles of axial loading with a maximum force of 120 N. For the pull-off testing, before and after the same cyclic loading, the required force for disconnecting the remaining 6 implant-abutment connections was measured. The surface of 10 abutments was examined using a scanning electron microscope 120× before and after loading. Results. The pull-off test showed a significant decrease in the vertical force required to pull the abutment from the implant with mean 229.39 N ± 18.23 before loading, and 204.30 N ± 13.51 after loading (P <.01). Apart from the appearance of polished surface areas and slight signs of wear, no visible damages were found on the abutments. Conclusion. The deformation on the polished abutment surface might represent the result of micro movements within the implant-abutment connection during loading. Although there was a decrease of the pull-off force values after cyclic loading, this might not have a notable effect on the clinical performance.

목차

등록된 정보가 없습니다.