    Intermittent catheterization (IC) utilizing conventional eyelets catheters (CECs) for bladder drainage has long been the standard of care. However, when the tissue of the lower urinary tract comes in close proximity to the eyelets, mucosal suction often occurs, resulting in microtrauma. This study investigates the impact of replacing conventional eyelets with a drainage zone featuring multiple micro-holes, distributing pressure over a larger area. Lower pressures limit the suction of surrounding tissue into these micro-holes, significantly reducing tissue microtrauma. Using an ex vivo model replicating the intra-abdominal pressure conditions of the bladder, the intra-catheter pressure was measured during drainage. When mucosal suction occurred, intra-catheter images were recorded. Subsequently affected tissue samples were investigated histologically. The negative pressure peaks caused by mucosal suction were found to be very high for the CECs, leading to exfoliation of the bladder urothelium and breakage of the urothelial barrier. However, a micro-hole zone catheter (MHZC) with a multi-eyelet drainage zone showed significantly lower pressure peaks, with over 4 times lower peak intensity, thus inducing far less extensive microtraumas. Limiting or even eliminating mucosal suction and resulting tissue microtrauma may contribute to safer catheterizations in vivo and increased patient comfort and compliance.