The Korea Research Institute of Standards and Science (KRISS) established a new standard of the absorbed dose to water in LINAC X-ray beams. To confirm the equivalence of the new standard with other national metrology institutes (NMIs), a bilateral comparison study of the absorbed dose to water in high energy X-ray beams was performed between the KRISS and the National Metrology Institute of Japan (NMIJ). The comparison was made in-directly. Three transfer chambers were calibrated in the high energy X-ray beams by both laboratories and the calibration coefficients were compared. The average ratios of the calibration coefficients of the three transfer chambers obtained by the KRISS to those obtained by the NMIJ were 1.004, 1.006, 1.006, 1.007 for 6, 10, 15 and 18 MV X-ray beams, respectively. The calibration coefficients obtained at the KRISS were higher than those at the NMIJ but they were in good agreement within the expanded uncertainty of 1.0% (k ¼ 2). The results of this study will be used as the evidence for the KRISS standard being comparable with those of other NMIs, temporarily, in the interim period up to finalizing a key comparison study, BIPM.RI(I)eK6 managed by the Consultative Committee for Ionizing Radiation