논문 기본 정보

자료유형
학술저널
저자정보
장기현 (서울대학교 의과대학 방사선과학교실)
저널정보
대한영상의학회 대한방사선의학회지 대한방사선의학회지 제21권 제1호
발행연도
1985.1
수록면
17 - 30 (14page)

이용수

표지
📌
연구주제
📖
연구배경
🔬
연구방법
🏆
연구결과
AI에게 요청하기
추천
검색

초록· 키워드

오류제보하기
Magnetic resonance (MR) imagings with 0.15 Tesla resistive magnet developed by Korea Advanced Institute of Science were performed in 27 patients with various neurologic deseases and compared with X-ray computed tomography (CT). The purpose of the paper is to evaluate th image quality, the diagnostic value and limitation, and the optimal pulse sequence of MR imagings with a resistive magnet. The MR images were obtained by using a variety of pulse sequence with spin echo technique including saturation recovery, T2-weighted spin echo, and/or inversion recovery with various pulse repetition(TR) and echo delay(TE) times. The MR imaging demonstrated the capability of detecting the lesions shown on CT in all cases and also detected an additional finding in one case (multiple sclerosis) which was not seen on CT. The MR imaging appeared to be more useful than CT in the evaluation of syringomyelia of spinal cord and white matter disease, while it failed to demonstrate small calcific les on or inflammatory nodule(less than 1 cm) shown on CT and has shown somewhat poor contrast resolution in the case of meningioma. The spatial resolution of saturation recovery images was similar or superior to CT, whereas the contrast resolution of satruation recovery was inferior to CT. While the saturation recovery images have shown false negative findings in 5 patients (19%), the inversion recovery and T2-weighted spin echo images demonstrated better contrast discrimination between normal and pathologic conditions than the saturation recovery images, but somewhat poorer spatial resolution. Authors suggest that the MR images of both the saturation recovery with 300/30 and T2-weighted spin echo with 1000/90 be used as a routine procedure and additional iversion recovery of 1300/300/30 sequence as a option if white matter disease is suspected.

목차

등록된 정보가 없습니다.

참고문헌 (0)

참고문헌 신청

함께 읽어보면 좋을 논문

논문 유사도에 따라 DBpia 가 추천하는 논문입니다. 함께 보면 좋을 연관 논문을 확인해보세요!

이 논문의 저자 정보

최근 본 자료

전체보기

댓글(0)

0